Flex Your BICEPS

The tool you need for placing subcuticular sutures quickly!


Use the BICEPS to streamline any surgical task that requires rapid alternation between toothed and non-toothed forceps.

BICEPS advantages:

- Saves Time:  Especially for layered closure of incisions and superficial lacerations.

- Maximizes  Efficiency:  Minimizes instrument exchanges.

- Improves Cosmetic Outcomes:  Facilitates placement of subcutaneous sutures, delivering superior aesthetic results compared to staples.

- Reduces Infection Risk:  Results in fewer instrument transfers between the surgical tray and the operating field.

- Versatile Design:  Easily separable into individual toothed and non-toothed forceps if so desired.

- Ergonomic Handling:  Features weight distribution toward the handle which balances the BICEPS out of the incision or laceration.

- Less autoclaving:  Sterilize one instrument instead of two.

BICEPS bona fides:

- US Patent 11771418B2

- FDA approved

- ISO 9001, ISO 13485, CE

- Medical grade stainless steel

- Designed in USA

- Made in Pakistan

Using the BICEPS aligns with your commitment to providing high-quality, innovative solutions for your patients!  Put down the stapler.  Flex your BICEPS!.

Instructions for Use

INSTRUCTIONS FOR USE (IFU)


BICEPS: a Combination Forceps - Stainless Steel Handheld Surgical Instrument

Class I Medical Device

Manufacturer/Specification Developer: Joe Fiore Enterprises LLC

Patented Design - US Patent No. 11,771,418 B2


1. Device Description

The BICEPS is a combination forceps. It is a reusable, handheld stainless steel surgical instrument featuring toothed prongs (for tissue grasping and positioning) on one side, and non-toothed prongs on the other side (for needle or suture manipulation). It is designed for efficient single-tool use in surgical procedures without the need to switch instruments.


2. Intended Use/Indications

The BICEPS is intended for use in surgical settings to grasp, hold, manipulate, and position tissue, organs, sutures, or needles during procedures. It is reusable and must be cleaned and sterilized before each use.


3. Contraindications

Do not use on patients with known hypersensitivity to stainless steel or its components. Not intended for single-use or as a replacement for specialized instruments where the design does not meet procedural requirements.


4. Warnings and Precautions

- The device is supplied NON-STERILE. It must be thoroughly cleaned, inspected, and sterilized before first use and after every subsequent use.

- Inspect for damage, corrosion, misalignment, or wear before each use. Do not use if defective.

- Handle individually; avoid stacking or allowing contact with dissimilar metals during processing to prevent corrosion or damage.

- Use only neutral pH (approximately 7) enzymatic detergents. Avoid harsh chemicals, bleach, or abrasive cleaners, as they can cause pitting, staining, or corrosion.

- Ensure proper lubrication after cleaning and before sterilization.

- Flash sterilization may reduce instrument lifespan and is not recommended for routine use.


5. Instructions for Cleaning and Decontamination (Point-of-Use and Manual/Automated)

Immediately after use (to prevent drying of residues):

a. Rinse under warm (not hot) running water to remove gross soil, blood, and debris.

b. Soak in a neutral pH enzymatic detergent solution (follow manufacturer instructions, typically 3-10 minutes). Fully submerge.

c. Manual Cleaning: Use a soft bristle brush or nylon brush to clean all surfaces. Rinse thoroughly with distilled or demineralized water.

d. Ultrasonic Cleaning (recommended): Place in a neutral pH solution. Cycle 5-15 minutes (per manufacturer guidelines). Rinse thoroughly.

e. Dry completely with lint-free cloth or air dry. Inspect for cleanliness and functionality.

f. Lubricate with instrument milk or approved surgical lubricant.

Automated washer-disinfectors may be used if validated and following device-compatible cycles. Change solutions frequently.

Disinfection (if needed prior to sterilization): Use EPA-approved solutions per manufacturer instructions (aldehyde-free preferred). Rinse thoroughly. Note: This does not achieve sterility.


6. Sterilization (Steam Autoclaving Only)

Steam sterilization (autoclaving) is the recommended method for stainless steel instruments. Recommended Cycles (wrapped instruments):

- Gravity Displacement: 121 °C (250°F) for 30 minutes, followed by 3 0-4 5 minutes drying.

- Pre-Vacuum (Prevac): 132°C (270°F) for 4 minutes, followed by 30 minutes drying.

(Alternative: 134°C / 273°F for 3 minutes where validated.)

- Wrap singly or insets using FDA-cleared sterilization pouches or wraps suitable for steam.

- Use biological and chemical indicators per facility protocols.

- Allow full drying cycle to prevent moisture-related corrosion.

- Do not exceed recommended temperatures or times. Avoid ETO, cold chemical, or plasma sterilization.


7. Inspection and Maintenance

- After sterilization and before use, visually inspect for: cleanliness, corrosion, pitting, cracks proper prong alignment/closure, smooth operation, and integrity of tips.

- Lubricate as needed.

- Store in a clean, dry, dust-free environment.

- Proper care extends service life. Damaged instruments must be discarded.


8. Storage and Handling

Store sterilized instruments in their sterile packaging in a cool, dry location until use.


9. Additional Information

- These instructions are based on standard practices for reusable stainless steel surgical instruments. Users should follow their facility's validated protocols and applicable standards (e.g., AAMI, AORN).

- Report any serious incidents to the manufacturer and relevant authorities.

- For questions, contact: joe@joefioreenterprises.com.

- Version/Date: Revision: 1 / Date: July 2026

- These Instructions for Use do not replace professional training.